PSA file: C:\Users\ctd006\Desktop\OUTPACE_CTD\CTDclassik\Seasave_OUTPACE_CTD.psa Date: 02/27/2015 Instrument configuration file: C:\OUTPACE_CTD_Classik_DATA\CTD001_29760_OUTPACE_20150227..xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 0970 Calibrated on : 12-Aug-14 G : 4.79760659e-003 H : 6.55738391e-004 I : 2.29633715e-005 J : 1.65762044e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 0606 Calibrated on : 13-Aug-14 G : -1.03338854e+001 H : 1.37339787e+000 I : -1.30847913e-003 J : 1.59438161e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 29760 Calibrated on : 20 mars 2012 (SHOM) C1 : -4.040084e+004 C2 : 7.828816e-001 C3 : 1.138376e-002 D1 : 3.339932e-002 D2 : 0.000000e+000 T1 : 2.913178e+001 T2 : -7.258046e-005 T3 : 4.802986e-006 T4 : 0.000000e+000 T5 : 0.000000e+000 Slope : 0.99868800 Offset : -6.90100 AD590M : 1.166000e-002 AD590B : -8.703800e+000 4) Frequency 3, Temperature, 2 Serial number : 1328 Calibrated on : 16-Oct-14 G : 4.80610912e-003 H : 6.63658002e-004 I : 2.44381772e-005 J : 1.92635770e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 1073 Calibrated on : 13-Aug-14 G : -4.08662903e+000 H : 5.91430379e-001 I : 6.18124149e-005 J : 3.20943967e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Oxygen, SBE 43 Serial number : 0514 Calibrated on : 30-Jul-14 Equation : Sea-Bird Soc : 4.68640e-001 Offset : -5.06400e-001 A : -3.35580e-003 B : 1.91820e-004 C : -2.99440e-006 E : 3.60000e-002 Tau20 : 1.30000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 7) A/D voltage 1, Free 8) A/D voltage 2, Fluorometer, Chelsea Aqua 3 Serial number : 088-081 Calibrated on : 09/05/2014 VB : 0.475200 V1 : 1.985700 Vacetone : 0.501800 Scale factor : 1.000000 Slope : 1.000000 Offset : 0.000000 9) A/D voltage 3, Transmissometer, WET Labs C-Star Serial number : 373DR Calibrated on : 20140213 M : 21.6720 B : -1.2170 Path length : 0.250 10) A/D voltage 4, User Polynomial Serial number : 003 Calibrated on : 2012/06 Sensor name : UVP5 A0 : 0.00000000 A1 : 413.40000000 A2 : 0.00000000 A3 : 0.00000000 11) A/D voltage 5, PAR/Irradiance, Biospherical/Licor Serial number : 70494 Calibrated on : 27/12/2012 M : 1.00000000 B : 0.00000000 Calibration constant : 19900000000.00000000 Multiplier : 1.00000000 Offset : -0.05107000 12) A/D voltage 6, Altimeter Serial number : 5301.205445 Calibrated on : 18/03/2009 Scale factor : 30.000 Offset : 0.000 13) A/D voltage 7, Free 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 20127 Calibrated on : 21/03/2014 Conversion factor : 1387.30000000 Ratio multiplier : 1.00000000 Scan length : 40 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: C:\OUTPACE_CTD_Classik_DATA\out_c_033.hex Timeout (seconds) at startup: 60 Timeout (seconds) between scans: 20 --------------------------------------------- Instrument port configuration: Port = COM1 Baud rate = 19200 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: SBE Carousel Number of bottles: 24 Port: COM3 Enable remote firing: NO Firing sequence: Sequential Tone for bottle fire confirmation uses PC sound card. --------------------------------------------- Header information: Header Choice = Prompt for Header Information --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth, Average Sound Velocity, and TEOS-10 Latitude when NMEA is not available: 0.00000000 Longitude when NMEA is not available: 0.00000000 Average Sound Velocity Minimum pressure [db]: 20.00000000 Minimum salinity [psu]: 20.00000000 Pressure window size [db]: 20.00000000 Time window size [s]: 60.00000000 Descent and Acceleration Window size [s]: 2.00000000 Plume Anomaly Theta-B: 0.00000000 Salinity-B 0.00000000 Theta-Z / Salinity-Z 0.00000000 Reference pressure [db] 0.00000000 Oxygen Window size [s]: 2.00000000 Apply hysteresis correction: 1 Apply Tau correction: 1 Potential Temperature Anomaly A0: 0.00000000 A1: 0.00000000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: NO --------------------------------------------- Mark Variables: Variables: Digits Variable Name [units] ------ --------------------- 0 Scan Count 4 Depth [salt water, m] 7 Conductivity [S/m] 5 Salinity, Practical [PSU] --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: NO XML wrapper and settings: NO Seconds between raw data updates: 0.00000000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC sound card. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO