Dissolved oxygen concentrations
(Miléna Bégovic, LPCM, Villefranche-sur-Mer, F 06238)
Dissolved oxygen concentrations were measured on samples using an automated Winkler titration system, and then used for the calibration of the CTD oxygen signal.
Winkler oxygen titrations are reported in the file O2_Winkler.xls, where the first column is the number of the CTD (PRO_xxx), with the corresponding number of the bottle (the sample depth). Concentrations are expressed in ml/l in the second column.
Reference :
Murray C.N. and J.P. Riley (1969) The solubility of gases in distilled
water and sea water- II. Oxygen. Deep-Sea Research, 16, 311-320.
NOTA BENE ()
L’oxygène est dosé selon la méthode classique de Winkler, dans les conditions données par Murray and Riley (1969).
Les flacons utilisés sont des flacons de DBO Wheaton de 60 ml.
Les réactifs, chlorure de manganèse et iodure de potassium + soude, sont ajoutés à l’échantillon d’eau de mer immédiatement après prélèvement.
L’acidification est faite au moment du dosage qui a lieu généralement le lendemain ou au maximum 3 ou 4 jours après le prélèvement.
Le dosage est fait directement dans les flacons dont le volume a été préétalonné. Le dosage par du thiosulfate environ 0.01 N (étalonné avec une solution d’iodate de potassium 0.01 N) est piloté par ordinateur, le point équivalent étant déterminé par un changement de potentiel entre deux électrodes de platine identiques.
Suivant Murray and Riley, le blanc des réactifs est estimé à 0.01 ml O2/l.
Miléna BEGOVIC
Laboratoire de Physique et Chimie
Marines